Sterilization Process is 21% of the scored CRCST exam. Answer 10 exam style questions on hydrogen peroxide and low-temperature sterilization and see exactly which concepts would cost you points on test day.
Start the 10 questions →A technician plans to wrap a heat sensitive camera head in a cotton huck towel and then a nonwoven wrap for a hydrogen peroxide gas plasma cycle. What is the problem?
These questions cover hydrogen peroxide packaging, hydrogen peroxide pouches, dryness for hydrogen peroxide, liquids in low temperature systems and hydrogen peroxide BI. Each one shows the correct answer and why every other option is wrong, so a miss tells you exactly what to review before the CRCST.
Prefer to read first? Every question from the practice set above, with the correct answer and why each wrong option is wrong.
Answer: A. Hydrogen peroxide systems cannot process cellulose materials such as cotton, linen, or paper. Cellulose absorbs the hydrogen peroxide, which can cause the cycle to abort or leave the device without adequate sterilant contact.
Why not B: Towels are not required for EtO either. The issue is that cellulose is incompatible with hydrogen peroxide.
Why not C: Moisture also causes peroxide cycles to abort, so dampening the cotton makes the problem worse.
Why not D: The nonwoven wrap is the compatible part. The cotton towel is the item that must go.
Exam tactic: Peroxide hates cellulose: no paper, linen, or cotton.
Answer: D. Pouches for hydrogen peroxide systems must be made of synthetic, non-cellulose material, such as a polyolefin and plastic pouch labeled for that process. Paper pouches absorb the sterilant and are not compatible.
Why not A: Standard steam pouches have a paper side, and paper is cellulose that absorbs hydrogen peroxide.
Why not B: Paper is incompatible, and folding pouches is not an acceptable packaging practice.
Why not C: Double pouching does not fix incompatible material. A paper pouch remains cellulose.
Exam tactic: Peroxide pouch = synthetic web, no paper.
Answer: B. Items must be completely dry before a hydrogen peroxide cycle. If moisture is present on the objects, the required vacuum cannot be achieved and the cycle aborts.
Why not A: Dry items are what the system requires, so dryness does not cause an abort.
Why not C: Synthetic, non-cellulose wraps are the correct packaging for this process.
Why not D: Internal chemical indicators made for hydrogen peroxide are expected in each package.
Exam tactic: Wet load in peroxide = cycle abort.
Answer: C. Hydrogen peroxide gas systems cannot process liquids. Liquids interfere with the vacuum and with sterilant contact, and the system is designed only for clean, dry devices listed in their IFU.
Why not A: Opening the vial makes things worse, since the liquid would disrupt the cycle and absorb sterilant.
Why not B: Cycle length does not make a liquid compatible with this process.
Why not D: Linen is cellulose and also incompatible. The answer is no regardless of wrapping.
Exam tactic: No liquids in a peroxide chamber.
Answer: B. Hydrogen peroxide sterilizers are monitored with biological indicators containing Geobacillus stearothermophilus spores, the same organism used for steam.
Why not A: Bacillus atrophaeus is the BI organism for ethylene oxide and dry heat.
Why not C: Escherichia coli is a vegetative bacterium, far too easy to kill to challenge a sterilizer.
Why not D: Mycobacterium tuberculosis is used to rate disinfectants, not as a sterilizer BI.
Exam tactic: Peroxide shares the steam bug: stearothermophilus.
Answer: A. The byproducts of a hydrogen peroxide gas plasma cycle, water vapor and oxygen, are nontoxic. That is why no aeration phase is needed, which is a major advantage over ethylene oxide.
Why not B: Sterilant is not meant to remain in the package. The process leaves nontoxic byproducts.
Why not C: Packaging is not meant to absorb peroxide. Absorbent cellulose packaging is incompatible.
Why not D: The process is low temperature, not a steam cycle.
Exam tactic: Peroxide leaves water and oxygen. No aeration.
Answer: A. Trays and accessories for hydrogen peroxide must be validated for that process and free of cellulose. Paper liners or cotton towels absorb the sterilant, and some tray materials are not compatible, so the tray IFU must list the system.
Why not B: Paper liners are cellulose and absorb peroxide, so they are incompatible.
Why not C: Cotton is cellulose. It absorbs the sterilant and can abort the cycle.
Why not D: Tray materials and design affect sterilant contact, so the tray must be validated for the process.
Exam tactic: Peroxide tray: validated, no paper, no cotton.
Answer: D. Chemical indicators are designed for a specific sterilization process. A steam indicator does not react to hydrogen peroxide conditions, so it gives no meaningful information about the peroxide cycle.
Why not A: Size is not the issue. The indicator chemistry does not match the process.
Why not B: Steam CIs come in many types for production loads. The Bowie-Dick pack is only one specific test.
Why not C: Steam indicators do not release moisture. They simply do not respond to peroxide.
Exam tactic: Match the indicator to the sterilant.
Answer: A. Hydrogen peroxide cycles are short, roughly 16 to 75 minutes depending on the sterilizer model and cycle selected, and leave only water vapor and oxygen, so items are ready without aeration. EtO needs a long cycle plus hours of aeration and involves a toxic, carcinogenic gas.
Why not B: Hydrogen peroxide cannot process liquids or cellulose such as linen and paper.
Why not C: Lumen penetration is limited by length and diameter in peroxide systems, so it is not superior to all methods.
Why not D: Hydrogen peroxide sterilizers require routine biological monitoring like other sterilizers.
Exam tactic: Peroxide = fast and clean. EtO = slow and toxic.
Answer: B. Liquid chemical sterilant systems process unwrapped, immersed devices, so items come out wet and unpackaged. They are intended for just in time use and are transferred to the point of use and used promptly, not stored as sterile items.
Why not A: Unwrapped, wet devices cannot keep sterility in storage.
Why not C: Wrapping after the cycle exposes the device to contamination and does not create a sterile package.
Why not D: Drying on a counter and pouching afterward contaminates the device. It would need full reprocessing.
Exam tactic: Liquid sterilant = just in time, never stored.
The free 20 question diagnostic covers all 7 areas of the CRCST outline and shows the 3 fixes worth the most points.
Take the free diagnosticSee all 32 CRCST practice topics →